Free-breathing whole-heart multi-slice myocardial T1 mapping in 2 minutes

  • Rui Guo
  • , Xiaoying Cai
  • , Selcuk Kucukseymen
  • , Jennifer Rodriguez
  • , Amanda Paskavitz
  • , Patrick Pierce
  • , Beth Goddu
  • , Reza Nezafat*
  • *此作品的通讯作者
  • Harvard University
  • Siemens

科研成果: 期刊稿件文章同行评审

摘要

Purpose: To develop and validate a saturation-delay-inversion recovery preparation, slice tracking and multi-slice based sequence for measuring whole-heart native T1. Method: The proposed free-breathing sequence performs T1 mapping of multiple left-ventricular slices by slice-interleaved acquisition to collect 10 electrocardiogram-triggered single-shot slice-selective images for each slice. A saturation-delay-inversion recovery pulse is used for T1 preparation. Prospective slice tracking by the diaphragm navigator and retrospective registration are used to reduce through-plane and in-plane motion, respectively. The proposed sequence was validated in both phantom and human subjects (12 healthy subjects and 15 patients who were referred for a clinical cardiac MR exam) and compared with saturation recovery single-shot acquisition (SASHA) and modified Look-Locker inversion recovery (MOLLI). Results: Phantom T1 measured by the proposed sequence had excellent agreement (R2 = 0.99) with the ground-truth T1 and was insensitive to heart rate. In both healthy subjects and patients, the proposed sequence yielded nine left-ventricular T1 maps per volume in less than 2 minutes (healthy volunteers: 1.8 ± 0.4 minutes; patients: 1.9 ± 0.2 minutes). The average T1 of whole left ventricle for all healthy subjects and patients were 1560 ± 61 and 1535 ± 49 ms by SASHA, 1208 ± 42 and 1233 ± 56 ms by MOLLI5(3)3, and 1397 ± 34 and 1433 ± 56 ms by the proposed sequence, respectively. The corresponding coefficient of variation of T1 were 6.2 ± 1.4% and 5.8 ± 1.6%, 5.3 ± 1.1% and 5.1 ± 0.8%, and 4.9 ± 0.8% and 4.5 ± 0.8%, respectively. Conclusion: The proposed sequence enables quantification of whole heart T1 with good accuracy and precision in less than 2 minutes during free breathing.

源语言英语
页(从-至)89-102
页数14
期刊Magnetic Resonance in Medicine
85
1
DOI
出版状态已出版 - 1 1月 2021
已对外发布

联合国可持续发展目标

此成果有助于实现下列可持续发展目标:

  1. 可持续发展目标 3 - 良好健康与福祉
    可持续发展目标 3 良好健康与福祉

指纹

探究 'Free-breathing whole-heart multi-slice myocardial T1 mapping in 2 minutes' 的科研主题。它们共同构成独一无二的指纹。

引用此