摘要
Transcranial focused ultrasound (TcFUS) neuromodulation is hindered by skull-induced acoustic limitations. To enable synergistic multi-region brain stimulation, we designed non-coaxial multi-focal composite Fresnel acoustic lenses, including an overlapping Fresnel lens (OFL) and an alternating-segmented Fresnel lens (ASFL). These lenses convert planar ultrasound into multiple foci. Based on Fresnel theory, acoustic fields were analyzed via simulations and experiments, validating the generation of four non-coaxial foci (10/30 mm focal lengths) from a 1 MHz planar wave using both OFL and ASFL. The influence of lens parameters on focal pressure distribution was investigated, and morphology was quantified using a linear least-squares method. Significant differences in focal morphology and intensity between OFL and ASFL provide crucial guidance for optimizing multi-target TcFUS strategies.
| 源语言 | 英语 |
|---|---|
| 期刊论文编号 | 3299 |
| 期刊 | Sensors |
| 卷 | 25 |
| 期 | 11 |
| DOI | |
| 出版状态 | 已出版 - 6月 2025 |
| 已对外发布 | 是 |
学术指纹
探究 'Design and Structure of a Non-Coaxial Multi-Focal Composite Fresnel Acoustic Lens for Synergistic Ultrasound Stimulation of Multiple Brain Regions' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver