Artifact reduction in free-breathing, free-running myocardial perfusion imaging with interleaved non-selective RF excitations

  • Hassan Haji-Valizadeh
  • , Rui Guo
  • , Selcuk Kucukseymen
  • , Xiaoying Cai
  • , Jennifer Rodriguez
  • , Patrick Pierce
  • , Beth Goddu
  • , Warren Manning
  • , Reza Nezafat*
  • *此作品的通讯作者
  • Harvard University
  • Siemens

科研成果: 期刊稿件文章同行评审

摘要

Purpose: To reduce inflow and motion artifacts in free-breathing, free-running, steady-state spoiled gradient echo T1-weighted (SPGR) myocardial perfusion imaging. Method: Unsaturated spins from inflowing blood or out-of-plane motion cause flashing artifacts in free-running SPGR myocardial perfusion. During free-running SPGR, 1 non-selective RF excitation was added after every 3 slice-selective RF excitations to suppress inflow artifacts by forcing magnetization in neighboring regions to steady-state. Bloch simulations and phantom experiments were performed to evaluate the impact of the flip angle and non-selective RF frequency on inflowing spins and tissue contrast. Free-running perfusion with (n = 11) interleaved non-selective RF or without (n = 11) were studied in 22 subjects (age = 60.2 ± 14.3 years, 11 male). Perfusion images were graded on a 5-point Likert scale for conspicuity of wall enhancement, inflow/motion artifact, and streaking artifact and compared using Wilcoxon sum-rank testing. Result: Numeric simulation showed that 1 non-selective RF excitation applied after every 3 slice-selective RF excitations produced superior out-of-plane signal suppression compared to 1 non-selective RF excitation applied after every 6 or 9 slice-selective RF excitations. In vitro experiments showed that a 30° flip angle produced near-optimal myocardial contrast. In vivo experiments demonstrated that the addition of interleaved non-selective RF significantly (P <.01) improved conspicuity of wall enhancement (mean score = 4.4 vs. 3.2) and reduced inflow/motion (mean score = 4.5 vs. 2.5) and streaking (mean score = 3.9 vs. 2.4) artifacts. Conclusion: Non-selective RF excitations interleaved between slice-selective excitations can reduce image artifacts in free-breathing, ungated perfusion images. Further studies are warranted to assess the diagnostic accuracy of the proposed solution for evaluating myocardial ischemia.

源语言英语
页(从-至)954-963
页数10
期刊Magnetic Resonance in Medicine
86
2
DOI
出版状态已出版 - 8月 2021
已对外发布

指纹

探究 'Artifact reduction in free-breathing, free-running myocardial perfusion imaging with interleaved non-selective RF excitations' 的科研主题。它们共同构成独一无二的指纹。

引用此