摘要
Blood glucose monitoring is crucial for the early diagnosis and treatment of diabetes; however, traditional electrochemical biosensors face challenges in detecting glucose in whole blood. To address this, we developed a fouling-resistant molecularly imprinted electrochemical biosensor. The sensor utilizes a nickel-based metal-organic framework (MOF), Ni3(HITP)2 (HITP = 2,3,6,7,10,11-hexaiminotriphenylene), as the modifier, o-phenylenediamine (o-PD) as the monomer, and glucose as the template to prepare the molecularly imprinted polymer (MIP). A phase-transition bovine serum albumin (PTB) anti-fouling layer was added to the electrode, resulting in the construction of the PTB/MIP/Ni3(HITP)2-MOF/SPCE sensor. This sensor demonstrates excellent antifouling capability, selectivity, and sensitivity, with a detection range of 1 μM to 100 mM and a limit of detection (LOD) of 0.31 μM. It requires no pre-treatment and can directly detect glucose in whole blood, offering a simple and direct method for clinical monitoring.
| 源语言 | 英语 |
|---|---|
| 文章编号 | 138028 |
| 期刊 | Sensors and Actuators, B: Chemical |
| 卷 | 441 |
| DOI | |
| 出版状态 | 已出版 - 15 10月 2025 |
联合国可持续发展目标
此成果有助于实现下列可持续发展目标:
-
可持续发展目标 3 良好健康与福祉
指纹
探究 'A novel molecularly imprinted electrochemical biosensor based on Ni3(HITP)2-MOF and a novel anti-fouling material for the direct detection of glucose in whole blood' 的科研主题。它们共同构成独一无二的指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS