摘要
In pulsed power systems, pulsed currents with risetimes from nanosecond to microsecond can be effectively measured by self-integrating Rogowski coils. Appropriate design of the structure and the integrating resistor is crucial to the high-frequency response of a coil. In this paper, several novel designs of Rogowski coil's integrating resistors were proposed and tested. Experimental results showed that the optimized coil could response square waves with fronts of ∼1.5 ns and had a sensitivity of ∼0.75 V/kA. The maximal peak current was designed as 100 kA.
| 源语言 | 英语 |
|---|---|
| 文章编号 | 035114 |
| 期刊 | Review of Scientific Instruments |
| 卷 | 86 |
| 期 | 3 |
| DOI | |
| 出版状态 | 已出版 - 1 3月 2015 |
| 已对外发布 | 是 |