A mini mass spectrometer with a low noise Faraday detector

  • Yang Tang
  • , Qian Xu
  • , Dayu Li
  • , Wei Xu*
  • *此作品的通讯作者
  • Beijing Institute of Technology
  • Northeastern University China

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

An ion trap mass spectrometer is conventionally featured with an electron multiplier as its detector. However, an electron multiplier can typically work at pressures below 20 mTorr with a high voltage applied, which limits the further miniaturization of ion trap mass spectrometers. In this work, a low noise Faraday detector was developed and integrated in our miniature mass spectrometer instrument, and a post data processing method was applied to improve its performance. A limit of detection of 1 ng mL-1 was achieved, and quantitation performance and mass resolution were characterized. This technology could be useful in the further development of miniature mass spectrometers by increasing background pressures.

源语言英语
页(从-至)3892-3898
页数7
期刊Analyst
卷145
期11
DOI
出版状态已出版 - 7 6月 2020

学术指纹

探究 'A mini mass spectrometer with a low noise Faraday detector' 的科研主题。它们共同构成独一无二的学术指纹。