Abstract
We study noise reduction using modified compressive sensing optical coherence tomography. We show that averaged modified compressed sensing (CS) reconstruction achieves better image quality in terms of signal-to-noise ratio, local contrast, and contrast-to-noise ratio, compared to the classical averaging method while reducing the total amount of data required to reconstruct the images. The same is also true when compared with standard CS-based averaging method with the same amount of undersampled data.
Original language | English |
---|---|
Pages (from-to) | 4209-4211 |
Number of pages | 3 |
Journal | Optics Letters |
Volume | 37 |
Issue number | 20 |
DOIs | |
Publication status | Published - 15 Oct 2012 |
Externally published | Yes |