Abstract
The simulative lung cancer tissues under the approximate clinical condition were imaged using in-line holography method with 35 keV synchrotron radiation hard X-ray. The millimeter scale simulative cancer phantoms showed adequate contrast to lung tissues in our experiment. It demonstrates that in-line holography method with synchrotron radiation hard X-ray promises to be a potential sensitive method for the early detection of lung cancer. The image contrast, standard deviation (SD) and normalized standard deviation (NSD) of different areas were calculated. It shows that the traditional method of contrast calculation does not always give a convincible result in image judgment; a standard deviation map of image taken with a proper distance of sample to detector (DSD) will correspond well to the projecting image and supply effective assistance in diagnostic judgment.
Original language | English |
---|---|
Article number | C07002 |
Journal | Journal of Instrumentation |
Volume | 8 |
Issue number | 7 |
DOIs | |
Publication status | Published - Jul 2013 |
Externally published | Yes |
Keywords
- Inspection with x-rays
- Medicalimage reconstruction methods and algorithms, computer-aided so
- X-ray radiography and digital radiography (DR)