Skip to main navigation Skip to search Skip to main content

FUV imaging spectrometer based on crossed czerny-turner structure

  • Yan Wu
  • , Yi Tang*
  • , Guo Qiang Ni
  • , Ting Zhu Bai
  • , Gang Huang
  • *Corresponding author for this work
  • Beijing Institute of Technology

Research output: Contribution to journal › Article › peer-review

Abstract

This article introduced the needs and key technologies of the Far Ultraviolet (FUV) imaging spectrometer. The design method and spectral calibration were researched. The FUV imaging spectrometer was based on crossed Czerny-Turner optical structure with all reflective components in it. The initial structure was calculated based on the primary aberrations theory, and the optical system whose effective wavelength range is from 100 nm to 300 nm was simulated and optimized by the optical design software Zemax in order to advance the spectral resolution. The crossed Czerny-Turner spectrometer system was set up and the spectral calibration and imaging ability test were achieved. The measured spectral bandwidth is less than 1.67 nm.

Original languageEnglish
Pages (from-to)125-129
Number of pages5
JournalGuangdian Gongcheng/Opto-Electronic Engineering
Volume36
Issue number3
Publication statusPublished - Mar 2009

Keywords

  • Czerny-Turner system
  • Far ultraviolet
  • Imaging spectrometer
  • Space weather
  • Spectral calibration

Fingerprint

Dive into the research topics of 'FUV imaging spectrometer based on crossed czerny-turner structure'. Together they form a unique fingerprint.