Experimental study on beam parameter measurement system by using area array CCD

Research output: Contribution to journalArticlepeer-review

Abstract

The technical problems concerning the determination of the laser beam parameters by using area array charge coupled device (CCD) detector were studied. The beam parameters of Hermite-Gaussian modes generated from laser diode (LD) pumped CW Nd:YAG laser with 1064 nm and a lamp pumped Q-switched Nd:YAG laser with 10 ns pulse width were experimentally measured. Comparing the measured beam quality factor M2 with theoretical one, the measuring deviation of M2 was less than 5%. The measuring error of the divergence was also less than 5%. The influences of various factors on the measured beam parameters were discussed.

Original languageEnglish
Pages (from-to)993-996
Number of pages4
JournalZhongguo Jiguang/Chinese Journal of Lasers
Volume32
Issue number7
Publication statusPublished - Jul 2005

Keywords

  • Area array CCD
  • Beam parameter measurement
  • Laser physics
  • M-factor
  • Mode generator
  • Physicelectronics

Fingerprint

Dive into the research topics of 'Experimental study on beam parameter measurement system by using area array CCD'. Together they form a unique fingerprint.