Design of a new TEC-based accurate incubator and test of MAX1978

  • Hong Bo Li*
  • , Wei Qian Zhao
  • , Si Tian Gao
  • , Jian Jun Cui
  • , Jian Ding
  • , Hai Jie Zhang
  • , Qiang Li
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

A constant temperature control system is design based on the thermoelectric cooler. Thermocouple is used as a temperature measurement sensor, whose measurement uncertainty is 6.6mK used in narrow temperature interval. The main thermostat and auxiliary thermostat core is MAX1978.The design goal of the main thermostat is to provide ±0.01°C temperature stability and uniformity inner the nano-scale measurement environment without heat-source. Auxiliary thermostat is regulated to achieve a good temperature distribution inner the nano-scale measurement environment within a little heat-power. The control-performance of MAX1978 is tested, which presents that the temperature stability is better than 3m°C/h, the temperature resolution is better than 0.01°C and the relation between temperature and input of MAX1978 is linear.

Original languageEnglish
Pages (from-to)216-221
Number of pages6
JournalJiliang Xuebao/Acta Metrologica Sinica
Volume31
Issue number6 A
DOIs
Publication statusPublished - Dec 2010

Keywords

  • MAX1978
  • Metrology
  • Precision temperature control
  • TEC
  • Thermocouple
  • Traceability

Fingerprint