Skip to main navigation Skip to search Skip to main content

Crosstalk reduction in optically pumped magnetometers arrays for biomagnetic measurement

  • Tianbo Wu*
  • , Wei Xiao
  • , Xiang Peng
  • , Teng Wu*
  • , Hong Guo
  • *Corresponding author for this work
  • Peking University
  • Beijing Institute of Technology

Research output: Contribution to journalArticlepeer-review

Abstract

Magnetoencephalography based on the compact OPM (Optically Pumped Magnetometer) array offers a series of advantages, such as low cost and wearability, and is gradually being used in brain science research and the diagnosis and treatment of brain diseases, such as epilepsy and dementia. As the distance between parametric resonance spin-exchange relaxation-free OPM sensors decreases, the array experiences increased crosstalk effect. We design a low-crosstalk coil to suppress magnetic field leakage, thereby reducing the impact of crosstalk. We compare the field leakage of the low-crosstalk coil and the Helmholtz-like coil and validate the performance of the coil using a 2 × 2 OPM array.

Original languageEnglish
Article number085004
JournalReview of Scientific Instruments
Volume96
Issue number8
DOIs
Publication statusPublished - 1 Aug 2025
Externally publishedYes

Fingerprint

Dive into the research topics of 'Crosstalk reduction in optically pumped magnetometers arrays for biomagnetic measurement'. Together they form a unique fingerprint.

Cite this