Accurate temperature control system for LD pumped single frequency NPRO laser

Yan Zheng*, Ran Wang, Fuyong Yue, Mingwei Gao, Chunqing Gao

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

Based on the requirement of LD pumped highly stable single frequency NPRO laser, an accurate temperature control system is developed. By using program-controlled P and PI arithmetic, the temperatures of the LD and laser crystal are accurately controlled from -10°C to +70°C with an accuracy of ±0.01°C through driving the thermoelectric cooler (TEC). Using the temperature control system, stable single frequency laser output at 1645 nm is obtained in a LD pumped Er:YAG NPRO laser, with a relative wavelength stability of 8.32×10-7 in 30 minutes.

Original languageEnglish
Pages (from-to)526-529
Number of pages4
JournalGuangxue Jishu/Optical Technique
Volume39
Issue number6
Publication statusPublished - Nov 2013

Keywords

  • Accurate temperature control
  • PID
  • Single frequency lasers
  • Wavelength stability

Fingerprint