Abstract
In order to improve the measurement accuracy of Coriolis mass flowmeters, this paper investigates the possible application of a strong tracking filter during signal processing. Specifically, a novel phase difference measurement method, based on a strong tracking filter for Coriolis mass flowmeters, is proposed in order to overcome the problems associated with the extended Kalman filters, which cannot accurately track time-varying phase differences. Compared with the existing methods, the proposed method can continuously track phase difference variation with high precision. The proposed method is a parallel algorithm that does not need to predict signal frequency in advance; this can effectively eliminate the quadratic error associated with frequency estimation. The results of simulation and experiments verify the proposed method that has better estimation performance and enhances the anti-interference performance of phase difference estimation, which contributes to improving the measurement accuracy of Coriolis mass flowmeters.
| Original language | English |
|---|---|
| Article number | 075003 |
| Journal | Review of Scientific Instruments |
| Volume | 90 |
| Issue number | 7 |
| DOIs | |
| Publication status | Published - 1 Jul 2019 |
| Externally published | Yes |
Fingerprint
Dive into the research topics of 'A phase difference measurement method based on strong tracking filter for Coriolis mass flowmeter'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver